Evaluation of the TSI Aerosol Impactor 3306/3321 System Using a Redesigned Impactor Stage With Solution and Suspension Metered-dose Inhalers
QuickNav:
Search:  
 View PDF Version of this article   Citing Articles   Email This Article
 
Table of contents
Abstract   Introduction   Materials and Methods   Results and Discussion   Conclusion   Acknowledgments   References  

Harris JA, Stein SW, Myrdal PB. Evaluation of the TSI Aerosol Impactor 3306/3321 System Using a Redesigned Impactor Stage With Solution and Suspension Metered-dose Inhalers. AAPS PharmSciTech. 2006; 7(1): Article 20. DOI:  10.1208/pt070120

Evaluation of the TSI Aerosol Impactor 3306/3321 System Using a Redesigned Impactor Stage With Solution and Suspension Metered-dose Inhalers
Julie A. Harris,1 Stephen W. Stein,2 and Paul B. Myrdal1

1College of Pharmacy, University of Arizona, Tucson, AZ
23M Drug Delivery Systems, Early Pharmaceutics and Technology Laboratory, St Paul, MN

Correspondence to:
Julie A. Harris
Tel: (520) 626-3847
Fax: (520) 626-4063
Email: harris@pharmacy.arizona.edu

There is a Corrigendum associated with this document:  AAPS PharmSciTech. 2006;7(3):E81. DOI: 10.1208/pt070381c

Received: June 17, 2005; Accepted: December 5, 2005; Published: March 10, 2006

Abstract

The purpose of this research was to evaluate a redesigned impactor stage for the TSI Model 3306 Impactor Inlet with nozzles adjusted to obtain a target cut-point of 4.7 μm. It has been determined that the previous cut-point used in the Model 3306 was nominally closer to 4.14 μm, thus potentially impacting the characterization of aerosol mass. The reassessment of the Model 3306 was performed on 4 solution and 2 suspension metered-dose inhaler (MDI) formulations. The redesigned impactor stage resulted in a 5% to 6% increase in aerosol mass when compared with the previous impactor stage for the products Ventolin-HFA, Proventil-HFA, and 2 cyclosporin solution formulations with high ethanol concentrations (15% wt/wt). For the formulations with low ethanol concentrations (3% wt/wt), minimal differences were observed between the 2 cut-points. In addition, this study reevaluated the requirement of a vertical inlet extension length when using the TSI 3306/3321 system with the redesigned cut-point. It was shown that the use of a 20-cm extension provides mass and aerosol size distributions that are comparable to the Andersen 8-stage Cascade Impactor, for both solution and suspension MDIs. This work indicates that the TSI 3306/3321 system is suitable for preformulation studies of both suspension and solution MDI systems.

Keywords: aerosols, metered-dose inhalers, particle sizer, size distribution

Introduction

The aerodynamic diameter of a medicinal aerosol has a direct bearing on the ability of the aerosolized drug to deposit in the lungs and is thus an important parameter when developing metered-dose inhalers (MDIs). Analysis of the particle distribution from a pressurized MDI is typically performed using a multi-stage cascade impactor equipped with United States Pharmacopeia/European Pharmacopeia (USP/Ph Eur) induction port. The Andersen 8-stage Cascade Impactor (ACI) (Thermo Andersen, Smyrna, GA) is a standard compendial method for particle size assessment. Testing with the ACI allows for the mass and particle size distribution of the aerosolized drug to be determined directly through chemical analysis of each impaction stage.

While reliable, the ACI is a labor-intensive test method, especially for early product development. As a result, there is an ongoing interest in the development of more efficient techniques for aerosol characterization.1,2 Several researchers have evaluated laser diffraction as a potential method for characterizing aerosols. For continuously nebulized solutions, correlations between laser diffraction and cascade impaction can be obtained, provided environmental conditions are controlled.3-6 Relationships between laser diffraction and cascade impaction for MDI systems are more challenging owing to the dynamic liquid-vapor regime surrounding the aerosolized particles.7-9 Phase Doppler particle size analysis has also been evaluated for aqueous aerosols10; however, as with laser diffraction, neither method directly quantifies the drug present in the aerosol.

Recent interest has centered on the TSI 3306/3321 system (TSI Inc, Particle Instruments, Shoreview, MN), which represents an efficient tool for direct particle size measurement, while affording the ability to determine drug mass.1,2,11 The TSI Model 3321 aerodynamic particle sizer (APS) aerosol spectrometer, is a real-time aerodynamic particle size analyzer based on time-of-flight (TOF) analysis. This laser-based particle sizing instrument measures the mass median aerodynamic diameter (MMAD) of aerosol particles based on the particle’s velocities immediately downstream of a flow-accelerating nozzle.1 The Model 3321 can be integrated with the TSI Model 3306 single-stage impactor inlet. The use of a 4.7-μm impaction stage allows for the quantification of the fine particle mass of the aerosolized drug less than 4.7 μm (FPM4.7 μm).

In previous reports the use of different vertical inlet extension lengths (0, 20, 40 cm) were evaluated with the TSI 3306/3321 system for solution aerosols.2,12 The purpose of an extension is to allow a longer physical path for the droplets to travel before reaching the impactor jet, thus facilitating particle evaporation. The degree to which the extension length affects the droplet distribution depends on the formulation. Formulations with minimal or no semivolatile constituents will likely be unaffected. Those formulations containing significant semivolatile compositions are much more likely to be affected. These past studies demonstrated that the TSI 3306/3321 system can be used to obtain FPM4.7 μm values similar to the ACI. However, it has recently been determined that the original cut-point supplied in the Model 3306 design was smaller than the labeled 4.7 µm. As a result, the quantification of FPM4.7 μm may have been inaccurate, most notably for MDI formulations with particle size distributions approaching the cut-point.

The objective of this investigation was to quantify the differences between the 2 different cut-points used with the Model 3306 and to reevaluate the utility of inlet extensions when comparing FPF4.7 μm data to those obtained from ACI testing. In addition to varied solution formulations, this study demonstrates the feasibility of using the TSI 3306/3321 system for characterizing suspension MDIs.

Materials and Methods

Formulations

Two suspension products, Proventil-HFA (containing ethanol and surfactant) and Ventolin-HFA (containing no cosolvent or surfactant) were obtained from a commercial pharmacy. Four experimental solution MDI formulations made with varying cyclosporine (CSP) and ethanol concentrations were also evaluated and are displayed in Table 1.

Table 1. The 4 Cyclosporin Measured Dose Inhaler Solution Formulations Evaluated With the TSI 3306/3321 System and Andersen 8-stage Cascade Impactor


Formulation Cyclosporin
(% wt/wt)
Ethanol
(% wt/wt)
HFA-134a
(% wt/wt)

A 0.1 3 96.9
B 0.1 15 84.9
C 2 3 95
D 2 15 83



ACI

The ACI was operated at a flow rate of 28.3 L/min. The ACI contained stainless steel collection plates, which were coated for the suspension formulation Ventolin-HFA. A microfiber filter was placed after the bottom impaction stage. Each canister was shaken and then primed by actuating 3 times to waste. Each MDI was shaken prior to use and actuated 5 times, with the mouthpiece of the inhaler coupled on axis with the entry to the induction port (n = 3 vials). Flow through the ACI was maintained for an additional 30 seconds following the final actuation. The impactor was disassembled and the amount of drug (active pharmaceutical ingredient) was recovered quantitatively from the stem, actuator, inlet, collection plates, and filter. Samples were then assayed by high-performance liquid chromatography (HPLC)-UV spectrophotometry at a wavelength of 276 nm for albuterol sulfate and 210 nm for CSP. The HPLC system consisted of a Waters 2690 separation module (Waters, Milford, MA) coupled with a Waters 2487 dual wavelength absorbance detector. Sample analysis was performed by a reverse-phase HPLC assay, using a 150 × 4.6 mm, 5 μm Apollo Silica column for albuterol sulfate, and a 150 × 3.2 mm, 5 μm Alltech C18 platinum column for CSP. The mobile phase for CSP was 70% acetonitrile and 30% water at a flow rate of 1.0 mL/min. The mobile phase for albuterol sulfate was 82% acetonitrile and 18% 0.1% ammonium acetate (pH 3.0) at a flow rate of 0.8 mL/min.

TSI Model 3306/3321 System

The Model 3306 and Model 3321 were operated together. Aerosol presentation to the Model 3306 (Figure 1) is through a USP/Ph Eur inlet, at a flow rate of 28.3 L/min, and past a single-stage impactor. For this study, the target cut-point of the impactor stage was 4.7 µm, thereby allowing for the collection of a fine particle mass less than 4.7 µm (FPM4.7 μm) from the filter. The larger aerosol particles that do not flow through to the filter are collected on the impaction plate. Vertical extensions of 20 and 40 cm, when used, were inserted between the USP/Ph Eur inlet and the Model 3306. A small portion of the aerosol that penetrates through the USP/Ph Eur inlet (0.2% of the sample) is sampled isokinetically for size measurement via the Model 3321. The Model 3321 analyzes particles as they pass individually through the measurement zone, where their aerodynamic size is determined by TOF. For testing, each vial was shaken and then primed by actuating 3 times to waste. Five actuations were delivered for each test, and 3 tests were conducted for each vial. The mass median aerodynamic diameter (MMAD) was calculated using the TSI software, Aerosol Instrument Manager, Rev B (2002). The plate and filter were assayed via HPLC-UV, as described in the ACI section.

Figure 1 A schematic diagram of the Model 3306, where the aerosol enters through the USP throat inlet. Particles larger than 4.7 μm are impacted on the impactor plate (nonrespirable) and particles smaller than 4.7 μm are collected on the filter (respirable). Courtesy of TSI Inc.


Filter Percentage

In order to quantify the effects that cut-point and extension length differences have on the aerosol classification via the Model 3306, it is necessary to evaluate the relative changes in the plate and filter depositions. To this end the filter percentage (Filter%) can be defined as:
Filter % = ( Filter Filter + Plate ) × 100 (1)

where Filter represents the aerosol mass that passes the impaction plate (FPM4.7 μm) and the Plate represents the mass that is deposited on the impaction plate. The Filter% describes the percentage FPM4.7 μm of the aerosol that is presented to the impactor jets. This quantification allows for the performance of the impactor stage to be evaluated in isolation from the stem, actuator, or USP/Ph Eur inlet. This method is reasonable since the performance of the impactor stage should be independent of deposition in the stem, actuator, and USP/Ph inlet. The plate percentage is correspondingly 100% minus the Filter% and is not presented.

Fine Particle Fraction FPF4.7 μm

The FPF4.7 μm has historically been used as guidance for the therapeutic portion of the inhaled drug mass capable of reaching the airways of the respiratory tract. The FPF4.7 μm is defined as the FPM4.7 μm divided by the total amount of drug that exits the MDI mouthpiece. For the TSI 3306/3321 system, FPF4.7 μm can be determined by the following equation:
FPF 4.7 μ m TSI = ( FPM 4 .7 μ m USP/Ph Eur + extension(s) + plate + FPM 4.7 μ m ) × 100 (2)

where the USP/Ph Eur inlet and extension(s) represent the amount of drug deposited on each. In general, the amount quantifiable on the extensions has been found to be very small or negligible.12 In this study, detectable quantities of drug were found on only 24 of the 72 experiments (solution MDI formulations C and D) performed using the 20- or 40-cm extension on the TSI. The average was 1.75% of the total drug delivered.

Correspondingly, for the ACI, total drug recovered on stages 3 through 7 and filter is considered to be the FPM4.7 μm, giving the following equation:
FPF 4.7 μ m ACI = ( FPM 4 .7 μ m USP/Ph Eur + stage s 0 - 2 + FPM 4.7 μ m ) × 100 (3)

where stages 0 to 2 are stages 0, 1, and 2 of the ACI. The FPF4.7 μm was determined for the 4 solution formulations and 2 suspension products for both the TSI 3306/3321 system and the ACI.

Impactor Stage Cut-Point Calculations

For a given impactor stage, the particle collection efficiency is highly dependent on the Stokes’ number (Stk50) for that particular particle size. The collection efficiency curve for a given impactor stage is a function of the Stokes’ number and can be numerically or experimentally determined.13,14 The most descriptive value of an impactor stage is the Stokes’ number at which 50% of the particles impact and are collected, Stk50. The aerodynamic cut-point of a stage, da50, is the aerodynamic particle diameter at which Stk = Stk50.14 It is possible to solve for the aerodynamic cut-point, da50, of a stage using the following equation:
d a 50 = 9 S t k 50 η π D j 3 N 4 ρ p Q C c (4)

where η is the viscosity of the air; Dj is the jet diameter; N is the total number of jets on the impactor stage; ρp is the density of the particle; Q is the total flow rate through the impactor stage; and Cc is the Cunningham slip correction, which is a function of particle size. Equation 4 can be used to estimate the cut-point of various impactor stages as a function of nozzle diameter. In order to do this, it is first necessary to have an estimate of the Stk50 of the stage.15,16 The manufacturer’s reported cut-point and target hole diameter (new cut-point value) were used to estimate a Stk50 value of 0.3924 for the Model 3306. Next, the “old cut-point” aerodynamic diameter, da50, was estimated by substituting the calculated Stk50 number and the actual jet diameter of the old design impactor stage into Equation 4. The old cut-point for the Model 3306 was calculated to be ~4.14 μm.

Results and Discussion

The Influence of Impactor Stage Cut-Point on Filter%

The impact that the erroneous Model 3306 cut-point had on estimation of FPM4.7 μm was determined experimentally for both solution and suspension MDIs and is described in the following sections.

Solution Formulations

The Filter% for the various CSP solution formulations are shown in Figures 2 and 3. Figure 2 demonstrates that for low concentrations of drug and ethanol (A: 0.1% wt/wt CSP, 3% wt/wt ethanol) the average increase in Filter% for the new cut-point is a modest 0.57%, (P = .73) owing to the relatively small aerosol distribution produced from the formulation (MMAD = 0.93, geometric standard deviation [GSD] = 1.68, see Table 2). The influence that the inlet extension length had on the Filter% will be discussed later. However, for formulation B, (that has the same low drug concentration, but a higher ethanol content, 15% wt/wt) differences between cut-points are slightly more evident (P = .07) (Figure 2). While the residual particle distribution determined from the Model 3321 for formulation B (MMAD = 1.11μm, GSD = 1.61; see Table 2) is nearly the same as that of formulation A, it is evident that the aerosol passing the Model 3306 was not completely dry, resulting in a larger intermediate particle size distribution. Correspondingly, the new cut-point on the Model 3306 resulted in a 4.2% average increase in Filter% as compared with the old cut-point (whereas formulation A showed virtually no difference). Formulations C and D (Figure 3) had a higher nonvolatile concentration (2% wt/wt), which results in a larger particle size (MMAD of 2.66 and 3.05μm, respectively; see Table 2) and a lower Filter% in general, owing to the larger particle size distribution. Formulations C and D also show an increase in Filter% for the new cut-point with each extension length. These formulations were more sensitive to the exact cut-point of the Model 3306 since a greater portion of the aerosol mass was between 4.14 μm and 4.7 μm. The aerosol in this size range would likely have been collected on the impactor plate for the old (erroneous) cut-point but would likely not have been collected on the impactor plate when the new cut-point on the Model 3306 was used. The average increase in Filter% for the new cut-point for formulations C and D are 2.8% and 6.0%, respectively (P value 0.01, 0.002, respectively). Once again, the formulation with 15% ethanol had a lower Filter% than the 3% ethanol formulation. This result may be explained by the slightly increased residual particle MMAD (Table 2), along with the fact that the size of the droplets was possibly larger owing to incomplete evaporation of the ethanol as the droplets pass through the impactor stage.

Table 2. A Comparison of Size Distribution Measurements for 4 Cyclosporin Measured Dose Inhaler Solutions and 2 Suspensions Using the 20-cm Extension and New Cut-Point for the TSI 3306/3321 System Formulation*


Formulation ACI
MMAD (µm)
ACI
GSD
APS
MMAD (µm)
APS
GSD

A 0.88 ± 0.05 2.07 ± 0.09 0.93 ± 0.02 1.68 ± 0.17
B 1.07 ± 0.02 2.21 ± 0.1 1.11 ± 0.02 1.61 ± 0.01
C 2.39 ± 0.02 1.91 ± 0.02 2.66 ± 0.01 1.76 ± 0.03
D 2.81 ± 0.06 2.24 ± 0.05 3.05 ± 0.05 1.88 ± 0.01
Proventil-HFA 2.41 ± 0.35 1.56 ± 0.02 2.16 ± 0.13 1.89 ± 0.51
Ventolin-HFA 2.36 ± 0.06 1.54 ± 0.01 2.15 ± 0.05 1.51 ± 0.04


*ACI indicates Andersen 8-stage Cascade Impactor; MMAD, mass median aerodynamic diameter; APS, aerodynamic particle sizer; and GSD, geometric standard deviation.


Figure 2 Percentage fine particle mass of aerosol that enters the Model 3306 versus extension length for old and new cut-points using formulations (A) 0.1% cyclosporin with 3.0% ethanol (solid symbols), and (B) 0.1% cyclosporin with 15.0% ethanol (open symbols).


Figure 3 Percentage fine particle mass of aerosol that enters the Model 3306 versus extension length for old and new cut-points using formulations (C) 2% cyclosporin with 3% ethanol (solid symbols), and (D) 2% cyclosporin with 15% ethanol (open symbols).


In general, it is clear that the new cut-point consistently resulted in a higher Filter%, which is in alignment with the cut-point calculations. In addition, Figures 2 and 3 reveal an increase in Filter% with an increase in extension length, which will be discussed later.

Suspension Formulations

The effect of cut-point differences were also evaluated using suspension MDI formulations. The Filter% for Ventolin-HFA and Proventil-HFA are shown in Figures 4 and 5, respectively. The average increase in Filter% for the new cut-point using Ventolin-HFA was 5.5%, (P = .01), while the percentage increase for Proventil-HFA averaged 6% (P = .002). Based on the particle size distributions for both Ventolin-HFA and Proventil-HFA (MMAD = 2.16, GSD = 1.51 and MMAD = 2.26, GSD = 1.89, respectively), this trend is consistent with that observed for the solution formulations C and D. The Filter% differences between old and new cut-point for both solution and suspension MDI formulations illustrate that the old cut-point has a smaller diameter, which resulted in a modest underestimation of the Filter% measured by the Model 3306.

Figure 4 Percentage fine particle mass of aerosol that enters the Model 3306 versus extension length for old and new cut-points using Ventolin-HFA suspension formulation.


Figure 5 Percentage fine particle mass of aerosol that enters the Model 3306 versus extension length for old and new cut-points using Proventil-HFA suspension formulation.


Influence of Extension Length on Filter%

Solution Formulations

In addition to investigating cut-point differences, this study reevaluated the appropriate use of an inlet extension based on formulation content for solution and suspension MDIs with the new cut-point on the Model 3306. Previous published work has shown that an extension is necessary for solution MDIs with higher concentrations of ethanol.2,12 For the solution MDIs in this study, as the vertical extension length increased from 0 to 40 cm, the Filter% also increased (Figures 2 and 3). The additional extension lengths had minimal effect on the particle distribution for formulation A (low drug and ethanol concentration) with an observed Filter% increase of 2% from 0 to 20 cm and no additional increase from 20 to 40 cm (Figure 2). Formulation C, containing the same ethanol concentration as formulation A (3%) and higher drug concentration (2%), had a slightly higher Filter% increase (5%) when going from 0 to 40 cm (Figure 3). The increase in extension length had the greatest effect on formulations B and D, which both contained 15% ethanol. These formulations had a total increase of ~8% in Filter% as the vertical extension length increased from 0 to 40 cm.

Suspension Formulations

This work also examined commercial suspension formulations with varying ethanol concentrations using the different inlet extension lengths. The Filter% for Ventolin-HFA, which does not contain a semivolatile cosolvent, was not significantly affected by the use of extension lengths (Figure 4). However Proventil-HFA, which does contain ethanol in the formulation, had a significant increase in Filter% (10%) as a function of extension length (Figure 5). As was the case for solution MDIs, the degree to which the extension length affects the Filter% for suspension MDIs depends on the nonvolatile composition in the formulation. As the vertical extension length increased from 0 to 40 cm, Proventil-HFA showed a clear increase in Filter%, whereas Ventolin-HFA did not. These trends were seen for both the old and new cut-points.

Comparison of TSI 3306/3321 System with the ACI

The TSI 3306/3321 system, an attractive instrument for rapid screening in early product development, was compared with results obtained via the ACI. The cut-point of stage 2 on the ACI was assumed to be ~4.7 μm (the nominal cut-point as indicated by the manufacturer). The optimal way to determine the actual cut-point of any given stage is to determine the collection efficiency curve of the stage using monodisperse reference particles. However, this is very labor intensive and was not used for this investigation. The collection efficiency curve of the ACI has been previously described.17 The exact cut-point of any given ACI stage has been shown to vary depending on the exact dimensions of the nozzles on the stage.18

The redesigned cut-point, along with the 20-cm extension, was used to analyze the utility of the TSI 3306/3321 system compared with the ACI for both solution and suspension MDI formulations. The FPF4.7 μm and particle size results were analyzed in order to evaluate comparability of the 2 systems.

Fine Particle Fraction (FPF4.7 μm)

The FPF4.7 μm was determined for the 4 solution formulations and 2 suspension products for both the TSI 3306/3321 APS system and ACI. Figure 6 compares the FPF4.7 μm determined for each formulation from the ACI to that from the TSI 3306/3321 system using the 20-cm extension and the new cut-point. Only 1 (Formulation C) of the 4 solution formulations had FPF4.7 μm values that were statistically different than the TSI 3306/3321 system and the ACI (P = .01). However, the magnitude of difference was small for this formulation. The other 3 solution formulations showed no statistical difference between the TSI 3306/3321 system and the ACI (P > .01). One of the 2 suspension formulations, Proventil-HFA, showed no statistical difference between the TSI 3306/3321 system (P = .332) and the ACI. Yet, Ventolin-HFA showed a statistical difference (P = .006), though this may be owing to experimental error and cannot be explained. The data illustrate that the TSI 3306/3321 system can be used in place of the ACI to determine FPF4.7 μm for solution and suspension MDI formulations.

Figure 6 Characterization of fine particle fraction (FPF4.7 μm) particle distribution using the new cut-point and 20-cm extension for solution and suspension formulations on the TSI 3306/3321 system and ACI.


Particle Size

The MMAD and GSD, which represent the measures of central tendency and spread, respectively, were used to compare the size distribution data from the Model 3321 and the ACI. When a solution MDI is actuated, the formulation is atomized into fine droplets, each containing a constant ratio of drug, propellant, and any cosolvent or excipients contained in the formulation. On the other hand, when a suspension MDI is atomized, not every droplet is uniform with respect to drug ratio. In fact, some of the atomized droplets may contain only excipient and propellant. Particle-sizing technologies, such as the TOF used with the Model 3321, which are nonspecific to chemical composition of each particle, are thus susceptible to misrepresentation of the drug particle size distribution when sizing the aerosol from an MDI suspension.19 With these physical differences in mind, it would not have been surprising to observe differences in MMAD between the TSI 3306/3321 system and ACI for suspension MDI formulations. However, Table 2 shows that, based on the current data using 4 solution and 2 suspension MDIs, the MMAD values from the ACI correlate well with MMAD values from the TSI 3306/3321 system, when the new cut-point and 20-cm extension are used.

Particle size values on the Model 3321 for Proventil-HFA were expected to be smaller than those for Ventolin-HFA, since Proventil-HFA contains surfactant. Yet, Table 2 shows that the values are similar, which leads to the belief that the mass of the droplets that contain only surfactant in Proventil-HFA does not cause a significantly biased measurement of the MMAD for the Model 3321.

Overall, the TSI 3306/3321 system shows similar patterns of particle distribution compared to the ACI for each of the solution and suspension formulations tested and is sufficiently sensitive to detect increases in solution concentration. Therefore, the TSI 3306/3321 system shows potential for future early development studies and may be used as a particle-sizing tool for suspension and solution MDIs of varying drug and ethanol concentrations.

USP/Ph Eur Inlet

The concentration of CSP and ethanol in the solution MDI formulations appeared to have an affect on USP/Ph Eur inlet deposition. Figure 7 shows the influence of formulation on the delivery characteristics from the CSP MDIs by comparing FPF4.7 μm and USP/Ph Eur inlet percentage, according to Equation 5:
USP/Ph Eur Percent = ( USP/Ph Eur USP/Ph Eur + plate + filter ) × 100 (5)

Figure 7USP/Ph Eur inlet percentage and fine particle fraction (FPF4.7 μm) as a function of cyclosporin concentration using the TSI 3306/3321 system and new cut-point.


The formulations with 3% ethanol (A, C) were much more efficient at delivering drug than the formulations with 15% ethanol (B, D) and had lower USP/Ph Eur inlet percentage and higher FPF4.7 μm. This finding is consistent with previously published literature that shows that deposition in the USP/Ph Eur induction port is inversely proportional to ethanol concentration.20 The concentration of CSP also had an influence on the delivery characteristics. The formulations with 3% CSP had a lower FPF4.7 μm than the formulations with 0.1% CSP. This result was seen in 2 different situations. First, the formulations with 3.0% CSP had a larger MMAD (Table 2) and therefore had a higher fraction of aerosol larger than 4.7 μm. Second, the USP/Ph Eur inlet percentage increased when CSP concentration increased from 0.1% to 3.0%. This difference was statistically significant, where the P = .005 for 3% ethanol (A vs C) and P = .002 for 15% ethanol (B vs D). Additional research is needed to further explain this trend.

Conclusion

This study shows, for the first time, good agreement between the TSI 3306/3321 system and ACI for both solution and suspension MDI formulations when an adequate extension and cut-point are used. For those formulations containing higher ethanol concentrations, a 20-cm extension may be necessary on the TSI 3306/3321 system in order to allow more time for the droplets to evaporate and to obtain particle size and FPF4.7 μm measurements that are in close agreement with the USP/Ph Eur standard ACI. This work indicates that the TSI 3306/3321 system is suitable for preformulation studies of both suspension and solution MDI systems.

Acknowledgments

The authors would like to acknowledge TSI Inc (Shoreview, MN) for their financial support of the APS cut-point reevaluation.

References

1. Mitchell JP, Nagel MW, Wiersema KJ, Doyle CC.  Aerodynamic particle size analysis of aerosols from pressurized metered-dose inhalers: comparison of Andersen 8-stage cascade impactor, next generation pharmaceutical impactor, and model 3321 aerodynamic particle sizer aerosol spectrometer. AAPS PharmSciTech. 2003;4:E54.
PubMed  DOI: 10.1208/pt040454

2. Mogalian E, Myrdal PB.  Application of USP inlet extensions to the TSI impactor system 3306/3320 using HFA 227 based solution metered dose inhalers. Drug Dev Ind Pharm

3. Wachtel H, Zeigler J.  Improved assessment of inhaler device performance using laser diffraction. Respiratory Drug Delivery VIII. 2002;2:379-381.

4. Wachtel H, Ziegler J, inventors. Laser diffraction method for particle size distribution measurements in pharmaceutical aerosols. US patent 0142665. June 30, 2005.

5. Finlay WH, Stapleton KW.  Undersizing of droplets from a vented nebulizer caused by aerosol heating during transit through an Andersen impactor. J Aerosol Sci. 1999;30:105-109.
DOI: 10.1016/S0021-8502(98)00024-X

6. Ding JY, McVeety BD, Busick DR, Miller PR, Zimlich WC, Placke ME.  Correlation of particle size distribution measurements between optical and inertial impaction techniques using an ethanolic drug formulation for inhalation. Respiratory Drug Delivery VIII. 2002;2:355-358.

7. Holmes CE, Kippax PG, Newell HE, Southhall JP, Ward DJ. Simultaneous analysis of respirable aerosols via laser diffraction and cascade impaction. Drug Delivery to the Lungs XII. London, UK: The Aerosol Society; 2001:58-61. 

8. Smyth HDC, Hickey AJ.  Comparative particle size analysis of solution propellant driven metered dose inhalers using cascade impaction and laser diffraction. Respiratory Drug Delivery VIII. 2002;2:731-734.

9. Haynes A, Shaik MS, Krarup H, Singh M.  Evaluation of the Malvern Spraytec with inhalation cell for the measurement of particle size distribution from metered dose inhalers. J Pharm Sci. 2004;93:349-363.
PubMed  DOI: 10.1002/jps.10558

10. Corcoran TE, Hitron R, Humphrey W, Chigier N.  Optical measurement of nebulized sprays: a quantitative comparison of diffraction, phase Doppler interferometry, and time-of-flight analysis. J Aerosol Sci. 2000;31:35-50.
DOI: 10.1016/S0021-8502(99)00030-0

11. Stein SW, Gabrio BJ, Oberreit D, Hairston P, Myrdal PB, Beck TJ.  An evaluation of mass-weighted size distribution measurements with the model 3320 aerodynamic particle sizer. J Aerosol Med. 2003;16:107-119.
PubMed  DOI: 10.1089/089426803321919870

12. Myrdal PB, Stein SW, Mogalian E, Hoye WL, Gupta A.  Comparison of the TSI model 3306 impactor inlet with the Andersen cascade impactor: solution metered dose inhalers. Drug Dev Ind Pharm. 2004;30:859-868.
PubMed  DOI: 10.1081/DDC-200034575

13. Marple VA. A Fundamental Study of Inertial Impactors [thesis]. Minneapolis, MN: University of Minnesota, Particle Technology Laboratory; 1970. Publication 144. 

14. Hinds WC. Aerosol Technology. New York, NY: John Wiley & Sons; 1999. 

15. Rader DJ, Marple VA.  A study of the effects of anisokinetic sampling. Aerosol Sci Technol. 1988;8:283-299.

16. Hickey AJ, Dalby RN, Byron PR.  Effects of surfactants on aerosol powders in suspension: implications for airborne particle size. Int J Pharm. 1988;42:267-270.
DOI: 10.1016/0378-5173(88)90186-X

17. Vaughan NP.  The Andersen impactor: calibration, wall losses and numerical simulation. J Aerosol Sci. 1989;20:67-90.
DOI: 10.1016/0021-8502(89)90032-3

18. Stein SW, Olson BA.  Variability in size distribution measurements obtained using multiple Andersen mark II cascade impactors. Pharm Res. 1997;14:1718-1725.
PubMed  DOI: 10.1023/A:1012175612193

19. Stein SW, Myrdal PB.  A theoretical and experimental analysis of formulation and device parameters affecting solution MDI size distributions. J Pharm Sci. 2004;93:2158-2175.
PubMed  DOI: 10.1002/jps.20116

20. Gupta A, Stein SW, Myrdal PB.  Balancing ethanol cosolvent concentration with product performance in 134a-based pressurized metered dose inhalers. J Aerosol Med. 2003;16:167-174.
PubMed  DOI: 10.1089/089426803321919924

Other works citing this article: 0
Show Citing Articles

A publication of the American Association of Pharmaceutical Scientists
2107 Wilson Blvd., Suite 700, Arlington, Virginia, 22201, USA
703-243-2800, Fax: 703-243-9650, aaps@aaps.org
Copyright ©2006. All Rights Reserved. ISSN 1530-9932.
Legal Disclaimer