| Otsuka M, Kato F, Matsuda Y, Ozaki Y.
Comparative Determination of Polymorphs of Indomethacin in Powders
and Tablets by Chemometrical Near-Infrared Spectroscopy and X-ray Powder
Diffractometry.
AAPS PharmSciTech. 2003; 4(2): article 19.
| Makoto Otsuka,1
Fumie Kato,1
Yoshihisa Matsuda,1
and Yukihiro Ozaki2
1Department of Pharmaceutical Technology, Kobe Pharmaceutical University, Higashi-Nada, Kobe 658-8558, Japan 2Department of Chemistry, School of Science, Kawansei-Gauin University, Sanda, 669-1337, Japan
Correspondence to: Makoto Otsuka Tel: +81 78-441-7531 Fax: +81 78-441-7532 Email: m-otsuka@kobepharma-u.ac.jp | Submitted: November 27, 2003; Accepted: March 19, 2003; Published: May 20, 2003 | Keywords:
near-infrared spectroscopy, chemometrics, polymorph, indomethacin, x-ray powder diffractometry | The purpose of this research was to develop a rapid chemometrical
method based on near-infrared (NIR) spectroscopy to determine indomethacin
(IMC) polymorphic content in mixed pharmaceutical powder and tablets. Mixed
powder samples with known polymorphic contents of forms α and
γ were obtained from physical mixing of 50% of IMC standard
polymorphic sample and 50% of excipient mixed powder sample consisting
of lactose, corn starch, and hydroxypropylcellulose. The tablets were
obtained by compressing the mixed powder at 245 MPa. X-ray powder diffraction
profiles and NIR spectra were recorded for 6 kinds of standard materials
with various polymorphic contents. The principal component regression
analysis was performed based on normalized NIR spectra sets of mixed powder
standard samples and tablets. The relationships between the actual and
predicted polymorphic contents of form g in the mixed powder measured
using x-ray powder diffraction and NIR spectroscopy show a straight line
with a slope of 0.960 and 0.995, and correlation coefficient constants of
0.970 and 0.993, respectively. The predicted content values of unknown
samples by x-ray powder diffraction and NIR spectroscopy were reproducible
and in close agreement, but those by NIR spectroscopy had smaller SDs than
those by x-ray powder diffraction. The results suggest that NIR spectroscopy
provides a more accurate quantitative analysis of polymorphic content in
pharmaceutical mixed powder and tablets than does conventional x-ray powder
diffractometry.  |
Analytical methods for polymorphism1-3 in pharmaceutical bulk
powders include x-ray powder diffraction,4 differential scanning calorimetry
(DSC),5 thermogravimetric analysis, microcalorimetry,6 infrared spectroscopy,7 Raman spectroscopy,8 and dissolution kinetics.9 However, these methods are time-consuming and make it costly to maintain
high-quality preparations of samples and/or their measurements. In contrast,
near-infrared (NIR) spectroscopy is simple because of its method of
nondestructive sample preparation. Consequently, NIR spectroscopy is fast
becoming an important technique for pharmaceutical analysis. In addition,
chemometrics provides an ideal means of extracting quantitative information
from all chemical analytical data10,11 of multicomponent samples. Calibration
methods such as multiple linear regression (MLR), principal component
analysis/principal component regression (PCA/PCR), and partial least squares
regression are commonly used for quantitative NIR spectroscopy.12-20 In our previous studies,12,21 quantitative NIR spectroscopy was applied to
evaluate the quality of bulk powder of pharmaceutical products. The degree of
crystallinity and polymorphic content of indomethacin (IMC) bulk powder were
evaluated by using MLR and PCR. Those valued by NIR spectroscopy were consistent
with those obtained by the conventional x-ray powder diffraction, and the
accuracy of NIR spectroscopy was more than 50% higher than that of the x-ray
method. These results suggested that the quantitative method based on NIR was
more accurate and reproducible than x-ray powder diffraction and that handling
was easier with NIR than with the conventional methods. On the other hand, Blanco
and Villar22 and Patel et al23 reported that the polymorphic content in
the product consists of drug and excipients, because practical pharmaceutical
preparations consist of the bulk drug powder and various kinds of additive powders.
In their studies, a powder form sample was applied for the assay, but a tablet
form was not. Therefore, the purpose of this study was the systematic quantitative
determination of IMC polymorphs in final pharmaceutical preparations, both of
mixed powder and of tablet form samples by PCR based on NIR spectroscopy. NIR
spectroscopy's accuracy and experimental advantages were compared with those
of conventional x-ray powder diffraction.
 | | Materials The bulk powder of IMC was obtained from Yashiro Co (Tokyo, Japan).
Form a of IMC was prepared by using the following method: 10 g of IMC bulk
powder was dissolved in 10 mL of ethanol at 80°C. The undissolved
drug was filtered off. Then 20 mL of distilled water at room temperature was
added to the IMC-saturated ethanol solution at 80°C. The precipitated
crystals were removed by filtration using a glass funnel and then dried under
a vacuum at room temperature. Form γ of IMC was prepared by
recrystallization from ethyl ether at room temperature.5 Both sample
powders were sieved using 52- and 125-µm screens in a Row-tap shaker
(Tanaka Tech Co, Osaka, Japan). Crystalline α-lactose monohydrate
(Pharmatose 200M, DeMelkindustrie Veghel Co, Veghel, The Netherlands) and
cornstarch (Matsuya Chem Co, Tokyo, Japan) were used as a diluent and a
disintegrator, respectively. Hydroxypropylcellulose (HPC) (HPC-LH31; Nihon
Soda Co, Tokyo, Japan) was used as a binder. All other chemicals were of
analytical grade. Standard Sample PreparationThe excipient mixed powder was obtained by mixing lactose (DCL 21) 7 g,
cornstarch 30 g, and HPC (LH 31) 5 g in a mortar using a pestle. The mixed
powders were put into a 200-mL plastic bag, the bag was expanded by air, and
then the bag was closed using a rubber band. The bag was placed in the
chamber of a V-type mixer (Tokujyu Ind Co, Model V-1, capacity: 2 L, mixing
speed: 28 rpm) and mixed for 1 hour. Known quantities of standard IMC mixtures
(total amount was 10 g) were obtained by physical mixing of form a and form
γ IMC powders at various ratios (0, 20, 40, 60, 80, and 100 wt/wt%
form γ content) in a mortar using a pestle. The standard IMC mixtures
were mixed in a V-type mixer for 1 hour as described above. Known polymorphic
content standard IMC/excipient mixtures (SIEM) were obtained from the careful
physical mixing of 50% of the SIEM and 50% of the excipient mixed sample powder
in a V-type mixer for 1 hour as described above. Uniformity of the mixing
method was tested by measuring the drug concentration of the IMC γ
form and excipient mixed powder as follows: After IMC γ form (5 g)
and excipient mixed powder (5 g) were mixed using the above mixing method for
10, 20, 30, 60, and 90 minutes, 6 sample powders corrected randomly, and
measured individually IMC concentration by UV/visible spectrometer (160A,
Shimadzu Co, Kyoto, Japan). The coefficient of variation of the samples reduced
with an increase in mixing time, and the values were almost constant at 5%
after more than 60 minutes of mixing. Therefore, subsequently, all the samples
were mixed for 60 minutes by a V-type mixer. Tableting Compression ProcessA compression/tension tester (Autograph, model IS-5000, Shimadzu Co)
was used at 25 ± 1°C. The tablet sample (SIEMT) was obtained
to compression 300 mg of SIEM by 8-mm diameter punch with a flat surface in a
die at 245 MPa (maximum upper punch pressure, compression speed of 15 mm/min). X-ray Powder Diffraction AnalysisThe x-ray powder diffraction profiles were measured using the following
method. The SIEMT was carefully deaggregated in an agate mortar using a pestle,
to obtain the powder form of a ground tablet. The x-ray powder diffraction
profiles were obtained using an x-ray diffractometer (XD-3A, Shimadzu Co, Kyoto
Japan). The measurement conditions included (1) scan mode—step scan,
(2) target—Cu, (3) filter—Ni, (4) voltage—20 kV,
(5) current—20 mA, (6) receiving slit—0.1 mm, (7) time
constant—1 second, and (8) scan width—0.1 degree/step. The
silicon powder was measured before and after measuring the x-ray powder
diffraction profile of the sample, as an external standard for correction of
direct beam intensity. About 80 mg of each sample powder was carefully loaded
into a glass holder, and the sample surface was flattened softly without
particle orientation by using a spatula and a glass plate. After the x-ray
powder diffraction profiles of samples were measured under the above conditions,
the intensity values were normalized against the intensity of silicon powder
(2θ = 28.8°), which was the external standard. The calibration
curves for quantification of crystal content were based on the total relative
intensity of the highest 4 independent diffraction peaks, 2θ = 11.6,
19.6, 21.8, and 26.6°, of the form γ crystal. All data were
reported as the average of 5 runs. Thermal AnalysisDSC was performed with a Type 3100 instrument (Mac Science Co, Tokyo Japan).
The operating conditions in a open-pan system were as follows: (1) sample
weight—5 mg, (2) heating rate—10°C
min–1, and (3)
N2 gas flow rate—30 mL
min–1. NIR SpectroscopyNIR spectra were taken using a Fourier-transformed NIR spectrometer
(InfraProver, BRAN+LUEBBE Co, Norderstedt, Germany). Briefly, a fiber-optic
probe was inserted into the sample powder (2 g) in a 20-mL glass bottle. The
tablet samples were put on the probe and covered with aluminum foil to prevent
stray light from getting in. Their NIR spectra were measured. The top surface
of the fiber-optic probe had a 12-degree cut angle to prevent reflecting light.
The NIR spectra consisted of 459 data points between 1000 and 2222 nm, and 5
scans per sample were recorded. A ceramic (Coor's Standard) reference scan was
taken for each set of samples. NIR spectra of 6 kinds of SIEM and SIEMT were
recorded 5 times with the NIR spectrometer. A total of 30 spectra, consisting
of 24 spectra for the calibration set and 6 spectra for the prediction of
calibration, were analyzed by a PCR program associated with the SESAMI software
(BRAN+LUEBBE Co). Quantitative Analysis of Unknown SamplesUnknown polymorphic content IMC samples (UIM) were obtained by a metastable
a form transformation to γ form in ethanol reported earlier.5 The pure α form of IMC (5 g) was dissolved in 50 mL of ethanol in a 100-mL
glass beaker. The temperature was maintained in a water bath at 37°C.
The IMC suspension samples were withdrawn at appropriate time intervals, filtered,
and dried under a vacuum at room temperature.5 The UIM were sieved using 52-
and 125-µm screens as described above. The mixture of UIM and excipient
(UIEM) was obtained by mixing the UIM with 50% of the excipient mixed powder, and
the tablet (UIEMT) of UIEM was compressed as described above. The contents of
α and γ forms were determined by using x-ray powder diffraction
and NIR spectroscopy. All data were reported as the average of 5 runs, respectively.  | Characterization of Form α and Form γ IMC Figure 1 shows the x-ray powder diffraction profiles of the polymorphic
forms of IMC, additives, and SIEM. The main x-ray powder diffraction peaks of
the form γ (Figure 1A) were at 11.6, 16.8, 19.6, 21.8, and 26.6°
(2θ), and those of the form a (Figure 1B) were at 8.4, 14.4, 18.5, and 22.0° (2θ), as
reported previously.5 The diffraction profile
of form α was broader than that of form γ, which might be
related to the lower crystallinity of form α. The lactose (Figure 1E) showed x-ray powder diffraction patterns due to crystalline α-lactose
monohydrate, but cornstarch (Figure 1F) and HPC (Figure 1G) showed halo patterns
due to noncrystalline solids. The x-ray powder diffraction profile of the SIEM
with 0% γ form content (Figure 1C) showed peaks at 8.4, 14.4, 18.5, and
22.0° (2θ) due to form α, and those of the SIEM with
100% γ form content (Figure 1D) showed peaks at 11.6, 16.8, 19.6, 21.8,
and 26.6° (2θ) due to form γ. The results suggested
that polymorphic determination was possible by conventional x-ray powder
diffraction in the mixed pharmaceutical preparations.
Figure 2 shows the DSC profiles of the pure form α and form
γ of IMC. These DSC curves showed corresponding endothermic peaks at
155 and 162°C, respectively, which are attributable to sample melting,
as reported previously.21 These results suggested that the form α
and form γ of IMC used in the present study were highly pure. Form
α was enantiotropic form, but form γ was monotropic form.
Measurement of IMC Polymorphic Content in the SIEM and SIEMT by Conventional X-ray Powder Diffractometry The x-ray powder diffraction profiles showed 2 main causes of fluctuation
in the determination of crystal content: (1) intensity fluctuation of the x-ray
direct beam during measurement, and (2) crystal orientation when the sample powder
was loaded in the sample holder. To avoid fluctuation of direct beam intensity,
the peak at 2θ = 28.8° of silicon powder was measured as an
external standard for correction of the value of crystalline content. The calibration
curve for measuring the SIEM with 0% to 100% γ form content by conventional
x-ray powder diffraction was based on the total intensity of the 4 specific
diffraction peaks, 2θ = 11.6, 19.6, 21.8, and 26.6°. The 4
diffraction peaks with high intensity were measured to minimize systematic error
due to crystal orientation. The obtained calibration curve to evaluate the polymorphic
content of form α and γ ratio IMC in the SIEM measured using x-ray
powder diffraction and the plot of total peak high intensity of 4 specific peaks
against the form γ concentration was linear, with a slope of 0.011, an
intercept of 0.028, and a correlation coefficient constant of 0.986 by the least-squares
method. The polymorphic content in the SIEM was calculated based on this calibration
curve. Figure 3 shows a plot of the relation between the actual and predicted
polymorphic contents of the SIEM measured using x-ray powder diffraction. This plot
shows a linear relation. It has a slope of 0.960, an intercept of 0.011, and a
correlation coefficient of 0.970. However, it has slightly higher 95% confidence
levels for the prediction of individual y values and 95% confidence intervals of
regression, indicating that x-ray powder diffraction has relatively low accuracy in
the determination of crystalline content.
Measurement of IMC Polymorphic Content in the SIEM and SIEMT by Chemoinfometric NIR Spectroscopy Figure 4A shows the NIR spectra of the α and γ forms of IMC.
The form α and form γ IMC showed significant NIR spectral peaks. The
NIR absorption peaks of IMC were assigned.21 The absorption peaks at 1130 and 1180
nm are associated with a second overtone of CH stretching bands for the HC=CH group
and the aromatic ring, respectively. The peak at 1400 nm is associated with a second
overtone of the CH stretching band. The absorption peaks at 1665, 1700, and 1730 nm
are associated with the first overtone of CH stretching of the CH=CH group, the second
overtone of the CH3 group, and
the first overtone of CH stretching bands of the
CH3 group. The peak at 2145
is associated with a combination tone of the NH and C=O groups' stretching bands,
respectively. The peaks at 1665, 1700, and 1730 nm, and the baseline at around 1800 nm,
of form γ were higher than those of form α, but the peak at 2200
nm was lower.
Figure 4B shows the NIR spectra of the excipients lactose, starch, and HPC.
All NIR spectra of excipients showed significant NIR spectral peaks and were significantly
different from those of form α and form γ IMC. Figures 5A and 5B show the NIR spectra of the SIEM with 0% to 100% γ
form content and the SIEMT with 0% to 100% γ form content. The peak intensities
at 1700, 1730, 1800, and 2200 nm of both spectra of the SIEM and the SIEMT decreased
with increasing of form γ content. Therefore, the NIR spectra data were
applied to PCR analysis to establish a quantitative model to evaluate the polymorphic
content in complex pharmaceutical preparations.
PCR is presented as regression of Y on selected principal components of X.
Properties of PCR are given together with a discussion on selection of eigenvectors.
Since PCR is useful to understanding the relationship between the objective parameters
and the principal component in the spectra, PCR was applied to the present study. A spectrum including n
spectral data can be seen as a point in an
n-dimensional space. In multivariate
data analysis, PCA/PCR of a spectral data matrix is a basic tool. PCA/PCR decomposes
X into a score matrix
T times a loading matrix
P' plus a residual matrix
E18:
This decomposition is particularly useful for converting
E to a few information plots
(score plots and loading plots) and for modeling the systematic structure in
E. The NIR spectra for SIEM and SIEMT were performed as a pretreatment to
minimize experimental error by using transformations of absorbance, peak high-normalized
absorbance, and second derivative. The best conditions were determined to minimize
the root mean squared error of prediction (SEP, Equation 2).
where yp
is property value of prediction,
yr is
property value of reference, and
n is number of spectra
in the calibration set. Table 1 shows that SEPs of the correlation curves were calculated
based on the 6 spectral data for training/test set corrected by 3
transformations. As a result, the minimum SEP value was calculated
from normalized NIR spectra based on the 3–principal component
model after normalization. The multiple correlation coefficient constant,
the standard error of estimate, and the SEP of calibration models including
3 principal components were calculated and are summarized in Table 1.
Figure 6 shows the plots of the calibration data for the SIEM and
SIEMT obtained by NIR spectroscopy based on the 3–principal
components model between the actual and predicted values, respectively. The
relationships between the actual and predicted polymorphic contents of
SIEM and SIEMT show a straight line with a slope of 0.995 and 0.997, an
intercept of 0.525 and –0.142, and a correlation coefficient constant
of 0.993 and 0.995, respectively; predicted values were reproducible and
had small SDs.
Figures 7 and 8 show loading vectors corresponding to the principal
component (PC) of SIEM and SIEMT, respectively. In the PC1 of SIEM, there
are positive peaks at 2187, 1934, 1711, and 1680 nm, and negative peaks at
1869, 1732, 1697, 1663, 1653, and 1185 nm. In PC2, there are positive peaks
at 1989, 1920, 1729, 1697, 1663, 1653, 1185, and 1128 nm, and negative peaks
at 2187, 2110, and 1581 nm. On the other hand, in the PC1 of SIEMT, there
are positive peaks at 2187, 2003, 1938, 1769, 1711, and 1680 nm, and a negative
peak at 2110 nm. In PC2, there are positive peaks at 2187 and 1680 nm, and
negative peaks at 2142, 1989, 1869, 1758, 1729, 1697, 1663, 1653, 1185, and
1128 nm. The SEPs of PC1, PC2, and PC3 of SIEM were 20.05, 6.843, and 3.602,
but those of SIEMT were 37.96, 9.797, and 3.491, respectively. The SEPs
decreasing of SIEM and SIEMT were more significant at PC1 and PC2 than that
between PC2 and PC3, respectively. This result indicated that PC1 loading
vector of SIEM was more effective on the calibration model than that of SIEMT,
and that PC2 of SIEM was less than that of SIEMT. The pattern of PC1 loading
vector of SIEM was similar to PC2 of SIEMT, except for the peak at 1934 nm.
On the other hand, the peak intensities at 2142, 1989, 1869, 1758, 1729,
1697, 1663, 1653, 1185, and 1128 nm decreased with decrease of form γ
content in original spectra of SIEM and SIEMT (Figures 5A and 5B), but those
at 1680 and 2187 nm increased. The peaks in the loading vector of SIEM and
SIEMT were reflecting a change of absorption peak intensity. The fact that
SEP decreased indicated that the PC1 loading vector was the most important
factor in the calibration model of SIEM, and PC2 was that of SIEMT. The
result suggested that the loading vectors reflected the spectral difference
between α and γ forms. The calibration model difference
between SIEM and SIEMT based on the loading vectors might be reflected in
powder structure change during tableting compression, such as porosity, mean
particle size, and distribution in the powder bed.
Since the purpose of this study was to compare the accuracy of
quantitative NIR spectroscopy to evaluate polymorphic content in the SIEM
with that of conventional x-ray powder diffraction, the mean accuracy
(Am)
and the mean precision
(Pm)
were determined by Equations 3 and 4, respectively:
where Xc
is predicted value of content of form γ IMC,
Xt is actual
value of content of form γ IMC, and
n is number of experiments. The
Am and
Pm for
the SIEM and SIEMT evaluated by NIR spectroscopy and x-ray powder diffraction are
summarized in Table 2. The confidence levels for the prediction of individual y
values for the SIEM and SIEMT by NIR spectroscopy were much narrower than for the
conventional x-ray method, but the result was consistent with the x-ray method.
These results indicate that NIR spectroscopy has good linearity
(shown in Figures 7 and 8) and good precision and accuracy (shown in Table 2).
On the other hand, the x-ray method has good accuracy (shown in Table 2), but its linearity and precision (shown in Figure 4) are not as good. Thus, these
assays are found to be superior for quantitative analysis of IMC polymorphs in
the SIEM and SIEMT.
Comparative Evaluation of UIEM and UIEMT by Chemometric NIR Spectroscopy and
Conventional X-ray Powder Diffraction To check adequacy of the calibration models obtained in the above section,
the polymorphic content of UIEM and UIEMT were evaluate by NIR spectroscopy and
x-ray powder diffraction. Figure 9 shows the plots of the predicted form
γ content in the UIEM and UIEMT measured by x-ray powder diffractometry
against those by NIR spectroscopy, respectively. The relationships between the
predicted polymorphic contents of UIEM and UIEMT measured by x-ray powder
diffraction and NIR spectroscopy show a straight line with a slope of 1.306 and
1.328, an intercept of –7.625 and –21.43, and a correlation
coefficient constant of 0.894 and 0.818, respectively. The slopes of the plots
of UIEM and UIEMT were slightly different from those of SIEM and SIEMT, and the
difference between UIEM and SIEM might be related to crystallinity or particle
size during crystalline transformation of UIM, since the x-ray powder diffraction
was affected by crystallinity and particle size. However, the line represents
a satisfactory correlation between the 2 predicted values of form γ
IMC content in the UIEM and UIEMT. Thus, NIR spectroscopy is an effective method
for quantitatively evaluating polymorph content in the UIEM and UIEMT.

| Quantitative evaluation of IMC polymorphs in practical pharmaceutical
systems, such as mixed powder and tablet systems, by NIR spectroscopy with the
chemometric method was proven to be superior to conventional x-ray powder
diffractometry. NIR spectroscopy could be of use as a validation method in the
pharmaceutical industry—both in line production and in evaluating final
pharmaceutical products. 
|
1.
FDA Papers. Guidelines: manufacturing and controls for INDs and NDAs. Pharm Tech Japan. 1985;1:835-850.
2.
Haleblian JK. Characterization of habits and crystalline modification of solids and their pharmaceutical applications. J Pharm Sci. 1975;64:1269-1288.
3.
Otsuka M, Matsuda Y. Polymorphism, pharmaceutical aspects. In: Swarbrick J, Boylan JC, eds. Encyclopedia of Pharmaceutical Technology. Vol 12. New York, NY: Marcel Dekker; 1995:305-326.
4.
Yoshino H, Hagiwara Y, Kobayashi S, Samejima M. Estimation of polymorphic transformation degree of pharmaceutical raw materials. Chem Pharm Bull. 1984;32:1523-1536.
5.
Kaneniwa N, Otsuka M, Hayashi T. Physicochemical characterization of indomethacin polymorphs and the transformation kinetics in ethanol. Chem Pharm Bull. 1985;33:3447-3455.
6.
Ahmed H, Buckton G, Rawlins DA. The use of isothermal microcalorimetry in the study of small degree of amorphous content of a hydrophobic powder. Int J Pharm. 1996;130:195-201.
7.
Black DB, Lovering EG. Estimation of the degree of crystallinity in digoxin by X-ray and infrared methods. J Pharm Pharmacol. 1977;29:684-687.
8.
Taylor LS, Zografi G. The quantitative analysis of crystallinity using FT-raman spectroscopy. Pharm Res. 1998;15:755-761.
9.
Berridge JC, Jones P, Roberts-McIntosh AS. Chemometrics in pharmaceutical analysis. J Pharm Biomed Anal. 1991;9:597-604.
10.
Edlund U, Grahn H. Multivariate data analysis of NMR data. J Pharm Biomed. Anal. 1991;9:655-658.
11.
Lincoln D, Fell AF, Anderson NH, England D. Assessment of chromatographic peak purity of drugs by multivariate analysis of diode-array and mass spectrometric data. J Pharm Biomed Anal. 1992;10:837-844.
12.
Otsuka M, Kato F, Matsuda Y. Comparative evaluation of the degree of indomethacin crystallinity by chemoinfometrical Fourier-transformed near-infrared spectroscopy and conventional powder X-ray diffractometry. PharmSci. 2000;2, Article 9.
13.
Morisseau KM, Rhodes CT. Near-infrared spectroscopy as a nondestructive alternative to conventional tablet hardness testing. Pharm Res. 1997;14:108-111.
14.
Drennen JK, Lodder RA. Nondestructive near-infrared analysis of intact tablets for determination of degradation products. J Pharm Sci. 1990;79:622-627.
15.
Buchanan BR, Baxter MA, Chen TS, Qin XZ, Robinson PA. Use of near-infrared spectroscopy to evaluate an active in a film-coated tablet. Pharm Res. 1996;13:616-621.
16.
Frake P, Gill I, Luscombe CN, Rudd DR, Waterhouse J, Jayasooriya UA. Near-infrared mass median particle size determination of lactose monohydrate, evaluating several chemometric approaches. Analyst. 1998;123:2043-2046.
17.
Martents H, Næs T. Multivariate Calibration. New York, NY: John Wiley & Sons; 1989.
18.
Norris T, Aldridge PK, Sekulic SS. Determination of end-points for polymorph conversions of crystalline organic compounds using on-line near-infrared spectroscopy. Analyst. 1997;122:549-552.
19.
Saver RW, Meulman PA, Bowerman DK, Havens JL. Factor analysis of infrared spectra for solid-state forms of delavirdine mesylate. Int J Pharm. 1998;167:105-120.
20.
Patel AD, Luner PE, Kemper MS. Low-level determination of polymorph composition in physical mixtures by near-infrared reflectance spectroscopy. J Pharm Sci. 2001;90:360-370.
21.
Otsuka M, Kato F, Matsuda Y. Determination of indomethacin polymorphic contents by chemometric near-infrared spectroscopy and conventional powder X-ray diffractometry. Analyst. 2001;126:1578-1582.
22.
Blanco M, Villar A. Polymorphic analysis of a pharmaceutical preparation by NIR spectroscopy. Analyst. 2000;125:2311-2314.
23.
Patel AD, Luner PE, Kemper MS. Quantitative analysis of polymorphs in binary and multi-component powder mixtures by near-infrared reflectance spectroscopy. Int J Pharm. 2000;206:63-74.

|
|