| Malik I, Poonacha M, Moses J, Lodder RA.
Multispectral Imaging of Tablets in Blister Packaging.
AAPS PharmSciTech. 2001; 2(2): article 9.
| Imran Malik,1
Mela Poonacha,1
Jennifer Moses,1
and Robert A. Lodder1
1Division of Pharmaceutical Sciences, College of Pharmacy, University of Kentucky Medical Center, Lexington, KY 40536-0082
Correspondence to: Robert A. Lodder Tel: 859-257-9232 Fax: Email: Lodder@pop.uky.edu | Submitted: February 27, 2001; Accepted: June 6, 2001; Published: June 19, 2001 | Keywords:
Near-infrared, Probability, Map, Aspirin, Moisture | This experiment tested the hypothesis that using near-infrared (IR) imaging
spectrometry on tablets through blister packs permits the identification and
composition of multiple individual tablets to be determined simultaneously.
Aspirin was selected for this study because its breakdown mechanism is well
understood. Near-IR cameras were used to collect thousands of spectra
simultaneously from a field of packaged aspirin tablets. Tablets were selected
by a principal component analysis selection algorithm. Graphs of the columns of
the transformation matrix showed that salicylic acid and acetylsalicylic acid in
the samples were modeled by the principal components. The bootstrap
error-adjusted single-sample technique chemometric-imaging algorithm was used to
draw probability-density contour plots that revealed tablet composition. Choice
of color was used to represent constituent identity, whereas intensity
represented concentration. The percentage of usable pixels in the indium antimonide
(InSb) array was 99.9%. The SEP was 0.06% of the tablet
mass for both water uptake and salicylic acid production. The number of tablets
that a typical near-IR camera can currently analyze simultaneously was also
estimated to be approximately 1300.  |
The perfect quality control method for pharmaceutical tablets would be
capable of rapid, nondestructive testing of intact tablets and capsules. This
testing would include determining multiple chemical constituents and physical
characteristics simultaneously. The method would be easy to use. Measurements
made with the technique would be free of systematic bias and highly
reproducible. The method would be selective, so no characteristic of the sample
or the environment would interfere with the measurement of any analyte or sample
property. As the instrument approaches the ideal "analytical black box," it
would be able to recognize that it is examining a sample unlike any it has ever
examined before and would respond appropriately. This response could take the
form of a request for operator assistance or for more samples of the same type,
a "second opinion" analysis by another technique, or a library search for the
best step to take. Near-infrared (near-IR) spectrometry with multivariate
chemometrics is becoming more and more like this ideal. The purpose of this
investigation was to test the hypothesis that near-IR imaging spectrometry
permits the identity and composition of large numbers of tablets to be
determined simultaneously through blister packs to study stability. Aspirin was
selected for study because its breakdown mechanism is well understood and
because it is inexpensive. Aspirin tablets can be placed in a blister pack so
that breakdown can be initiated easily by placing a small hole in the package
and putting the package into a hydrator. Near-IR spectrometry and nonparametric multivariate analysis are a
potent combination in solid dosage-form analysis, as demonstrated by 1. analysis of intact tablets,12. detection of tampering in gelatin capsules,2 and3. detection of contamination in drug capsules.3 Near-IR spectrometry and multivariate analysis has also been used to discriminate
between different tablet formulations inside blister packages.4 Near-IR spectrometry and multivariate analysis have even been employed to
determine the moisture and salicylic acid content of degraded aspirin tablets.5 Raman spectrometry with a near-IR light source has also been used on drug
formulations in gel capsules and from gel capsules inside blister packs.6 Analysis of the Raman spectra collected from bucindolol capsules inside blister
packs with multivariate calibration yielded a standard error of performance
(SEP) of only 3.36% of the range of active ingredient. As is often the case in
near-IR reflectance spectrometry, the largest source of error was the result of
sample inhomogeneity. Near-IR cameras are also being employed increasingly in multispectral imaging
experiments. Imaging spectrometers based on array cameras have fast scanning
ability and high sensitivity. Near-IR imaging was used in human stroke patients
to discover atherosclerotic plaque by identifying and locating oxidized
lipoprotein spectral signatures.7 The InSb camera employed had a custom cold (77oK) bandpass filter for near-IR
use and could be fitted with a warm (298oK) tunable interference filter system
or a warm filter disk. Probability density contours drawn in standard deviations (SDs) were used to form pictures that revealed the locations
of atherosclerotic plaque inside blood vessels. Near-IR multispectral imaging has been used to monitor solid-phase peptide synthesis.8 An acousto-optic tunable filter and a near-IR indium gallium arsenide (InGaAs) focal plane array camera maintained all the advantages
of a traditional near-IR spectrometer in noninvasive observation of reactions
and identification of the products during the solid-phase peptide synthesis. The
near-IR imaging system added a significant trait that traditional near-IR
spectrometers could not offer-the capacity to measure spectra at different
positions within a sample. In this study, spectra recorded by 16 × 16 pixels
were pooled to calculate a spectrum for each sample. Nevertheless, a good
spectrum could be collected from a single pixel. The kinetics of curing of an epoxy resin by amine was also studied using a
near-IR multispectral imaging spectrometer.9 The kinetics of curing was
calculated from data collected by a single pixel in the camera. The reaction
rates within the sample were not uniform. Because of this kinetic inhomogeneity,
differences in the degree of cure at different positions within the sample were
as high as 37% when data from only a single pixel were employed for calculation.
The inhomogeneity was not observed if the average of a large number of pixels
were used. In a similar manner, ethylene/vinyl acetate copolymers were shown to
exhibit a high degree of chemical inhomogeneity.10 This study tests the hypothesis that a near-IR camera and multispectral
imaging permits the identity and composition of large numbers of tablets to be
determined simultaneously through blister packs to study stability.
 | | Materials Near-IR images were obtained from 325-mg aspirin tablets
(Kroger, Cincinnati, OH). Reference spectra were obtained from blister packaging
(AH Robins, Richmond, VA), water, an aspirin tablet, and salicylic acid (Sigma,
St Louis, MO). Aluminum foil backing (Reynolds, Richmond, VA) and cement (DAP,
Dayton, OH) were used for sealing the polyvinyl chloride (PVC) blister packaging. Tetrabutyl ammonium phosphate (Eastman Kodak,
Rochester, NY), monosodium phosphate monohydrate (Fisher, Pittsburgh, PA),
ammonium hydroxide (Fisher), and methanol were used for the reference
high-performance liquid chromatography (HPLC) analysis of salicylic acid in aspirin. InstrumentationAn IRC-160 InSb focal plane array video camera
(Cincinnati Electronics, Mason, OH) with near-IR bandpass cold filter was used
for 0.5 m multispectral imaging of the samples. The camera frame rate was 51.44
frames/s and the photon energy response was 1800-10 000 cm-1. A polarizing
filter for the camera reduced specular reflectance (Polaroid, Cambridge, MA). A
tunable interference filter unit (OCLI, Santa Rosa, CA)11 with photon energy
response from 4325 to 5960 cm-1 was mounted on the camera. A Mitsubishi IR-M300
PtSi 256 X 256 CCD array camera (Cypress, CA USA) was used at 2 m on the samples. The light source was 2 250W
PC37771 near-IR/IR lamps (General Electric, Cleveland, OH). A monochromator
system employing a concave holographic grating (American Holographic, Fitchburg,
MA) and lead sulfide (PbS) detector was used for reference spectra.11 Whenever a near-IR camera was used to collect spectra, 2 spherical silicon
dioxide reflectance standards (1 high reflectance, 1 low reflectance) were
placed in each image to control for variations in light intensity and direction.
Images collected on different days and with the light sources in different
locations were made comparable by adjusting the gain and offset by
multiplicative scatter correction12 on the images so the intensities on the
standards were identical. The specular reflectance on the standards was used to
pinpoint the locations of the light sources and to provide a means to calibrate
reflected specular light intensity. Diffuse reflectance from the curved surfaces
of the 2 standards was used to calibrate shaded areas and sloping surfaces in
the images.13 Two light sources were employed for spectrometric imaging to
reduce shadows and achieve the best possible signal-to-noise ratio (S/N). The light sources were positioned at 90 degrees to each other with
the tablets at the vertex. The camera was between the light sources at 45
degrees to each. Spectral images were collected at each wavenumber with the
near-IR light sources off and again with them on to correct for the presence of
other lights in the room and for blackbody photon emission from the sample. A hydrator was constructed to control the conditions under which the tablets
decomposed.5 Tablets in blister packages were exposed to water vapor or a pH
9.0 ammonium hydroxide solution by punching a hole with a center punch through
the foil backing. Water absorption was determined by weighing and by near-IR
spectrometry. The HPLC for analysis of salicylic acid used a C-18 column (Analytical
Sciences, Santa Clara, CA), a Spectroflow 773 absorbance detector (ABI
Analytical, Ramsey, NJ), a model 6000A pump (Waters, Milford, MA), and a DATAQ DI-150 A/D computer interface (Akron, OH). The computer was an IBM-compatible personal computer. The
software for the near-IR equipment as well as the chromatography interface was
written in C++ (Microsoft, Redmond, WA) and Speakeasy (Speakeasy Computing Corp,
Chicago, IL). The mobile phase was 27:73 methanol:water, 0.15 M NaH2PO4, 5
mM tetrabutylammonium phosphate, at pH 6.0. This pH was selected to make the
results comparable with our earlier study.5 The half-life of aspirin at pH
6.0 is more than 5 days. Retention times for aspirin were approximately 5 minutes. Tablet masses were determined with an electronic balance. Tablets were
dissolved for HPLC analysis. The 43 tablets were selected from 172 hydrator
tablets by principal component analysis (PCA) selection algorithm.14 The advantage of this approach is that it selects the best tablets to use in the
calibration based on their near-IR spectra before reference measurements are
obtained. The near-IR method is fast and easy compared to the HPLC assay, making
the selection of the best subset of tablets for the reference assay based on
spectra worthwhile. The 172 tablets in turn came from 4 batches of 43 tablets,
each run for up to 24 hours in the hydrator.5 By staggering the starting
times of each 24-hour cycle it was possible to get at least 6 tablets in each
hour of exposure from 0 to 24 hours. The 4 runs were made over a period of 2 weeks. Imaging with the Bootstrap Error-adjusted Single-sample Technique (BEST).The BEST method was used to draw probability density contours for images at
various distances in SDs.13 In the BEST, a population P in a
hyperspace R represents the universe of possible spectrometric samples
(the rows of P are the individual samples, whereas the columns are the
independent information vectors, such as wavelengths or energies). P* is
a discrete realization of P based on a calibration set T, which
has the same dimensions as P* and is chosen only once from P to represent
as nearly as possible all the variations present in P. P* has parameters B and C, where C =
E(P) and B is the Monte Carlo approximation to the
bootstrap distribution. The expectation value, E( ), of P is the
center of P, and C is a row vector with as many elements as there
are columns in P. Each new sample spectrum X is projected into the hyperspace containing
B and many-one mapping the rows of B
onto the vector connecting C and X. X and C have
identical dimensions. The integral over the hyperspace R is calculated
from the center of P outward in all directions. The calculation of a
skew-adjusted SD is based on a comparison of the expectation value
C=E(P) and C=med(T), the median of T
in hyperspace (with the same dimensions as C) projected on the hyperline
connecting C and X. The result of the corrected projection is an
asymmetric SD that provides 2 measures of the standard deviation along the
hyperline connecting C and X.
Equation 1 defines the SD in the direction of X in hyperspace, and equation 2 defines the SD in the
opposite direction along the hyperline connecting C and X. Skew
adjusted SDs can be used to calculate mean distances between spectra of
different samples.  | Figure 1 shows representative spectra of an aspirin tablet (black line), the
blister package (red line), an aspirin tablet in a blister package (blue line),
and an aspirin tablet in a blister package with a leak exposing the tablet to
moisture (purple line). These spectra were obtained from single samples with the
nonimaging spectrometer. The major peaks of the packaging are at 5700 to 5800
cm-1 and 4000 to 4350 cm-1. The major spectral peaks of acetylsalicylic acid, at
about 6050 cm-1 and 4200 to 4800 cm-1, show up well through the packaging. The
most distinctive spectral feature of salicylic acid arising through aspirin
decomposition appears at 6370 cm-1.5 The major water-absorbance peak appears
at about 5200 cm-1. This region of the spectrum contains little in the way of
interference from the packaging, which simplifies the imaging of tablets.
Figures 2A and 2B show the calibration lines for water and salicylic
acid, respectively. In each case, the y-axis represents the result by the
reference method (weighing for water and HPLC for salicylic acid) whereas the
x-axis represents the result by multispectral imaging. The calibration lines
were constructed using data obtained at 0.5 m from 43 tablets selected from 172
by the PCA method of Svensson et al.14 The spectra of the cross-validation
samples (by leave-one-out method) are superimposed on the calibration line. The
standard error of estimate (SEE) for water was 0.05% of tablet mass and the SEP
was 0.06% of tablet mass. For salicylic acid, the SEE = 0.06% of tablet mass and
SEP = 0.06% of tablet mass.
Figure 3 reveals the spectra of water uptake by the 10 packaged tablets in the field of
packaged tablets shown in Figure 4. The tablets were
sampled at 8, 16, and 24 hours after a hole was punched in the foil backing. The
endpoint was at the same time for all tablets, and the starting time (ie, when
the hole was punched) was varied. These tablet spectra were collected through
the blister packaging, and are shown after multiplicative scatter correction.
The water peak appears again at 5200 cm-1 as a broad band beneath 3 other
smaller peaks. The peak on the left of the water band (5250 cm-1) arises from
the packaging material. The peak on the center of the water band (5200 cm-1)
arises from the aspirin, and the shifting peak on the right of the water band
(5150 cm-1) has overlapping components from both the aspirin and the blister
packaging. These spectra were used to calculate the contour
plot in Figure 4 and show that the changing signal from water over time is
readily detected through the blister packaging using a near-IR camera.
Figure 4 is a contour plot of tablets in blister packaging. The contours are
drawn in multidimensional standard deviations by the BEST method. A distance less than 3.8 SDs from a
calibration set of normal tablets is displayed as gray contour. A distance
greater than 3.8 SDs in the direction of the spectrum of water is colored red.
The intensity of the red color is correlated to the magnitude of the distance in
SDs between each pixel spectrum and the center of the spectra of control (dry)
tablets. The maximum distance is 8.7 BEST SDs (corresponding to the tablet with
the brightest red color in the contour plot, and highest spectrum at 5200 cm-1).
Previous studies in nonpharmaceutical applications have suggested that there
might be a minimum number of pixels required on a sample (eg, 16) to achieve an
acceptable S/N.8 Part of the S/N problem with focal plane arrays arises from inactive
("dead") pixels and "flickering" pixels. Some inactive pixels simply produce no output. However, inactive pixels often
produce a constant output outside the normal range of values from normal pixels.
The inactive pixel output value does not change with a changing optical signal.
Manufacturers frequently incorporate software corrections for these pixels into
their equipment. These corrections automatically replace the output values of
the inactive pixels with the output values of neighboring pixels. If a camera
performs such an inactive-pixel correction automatically on booting, it can
cause errors, especially when analyzing arrays of tablets in blister packaging.
When a large number of tablets are in the field of view, only a few pixels are
on each tablet; in these cases, it is easy for most or all of the values on an individual tablet
to be inaccurate. If the software cannot be bypassed, one
might never know that a tablet reading is essentially nonsense. For this reason,
access to raw data from the camera (as used in this study) is better than
corrected data. Flickering pixels have an output that varies randomly and somewhat
independently of the actual optical signal. This characteristic makes them much
harder to correct reliably with automatic routines in actual camera use because
the varying signal could be real. In addition, if the system averages frames to
produce a final image, simple averaging can obscure the flickering. Flickering
pixels are usually not corrected in camera software. Imaging reference cards or
scenes (in which the pixel output variation is known) during camera operation
will detect these pixels early, however. It should be noted that flickering
pixels often become inactive pixels with time. Approximately 0.1% of pixels in
an InSb focal plane array (FPA) are typically inactive or flickering, but
this number increases slowly with thermal cycling and age of the focal plane
array. Figure 4 suggests that maximum number of tablets that can be imaged
simultaneously on a 256 X 256 array is about 1300. With 1296 tablets in the camera field of
view, at least 16 pixels can be employed in simultaneous sampling of each
tablet, enough spectra on each tablet to get a useful S/N and draw contours for
imaging. Principal component analysis of the tablet spectra showed loadings with
spectral features known to correspond to salicylic acid and acetylsalicylic acid (Figure 5).5 Figure 5 shows the loadings column that correlated best to each
constituent. Acetylsalicylic acid features were noted on the loadings for
principal component 2, whereas salicylic acid features were observed on the
loadings for principal component 4.

| Multispectral imaging offers a potential speed advantage of approximately 30 000
over HPLC when determining moisture and salicylic acid in single, packaged aspirin tablets.
Although the imaging method is not as precise as spectrometric analysis of single tablets,
the precision is close.5 Multispectral imaging of a field of
tablets is approximately 1000 times faster than spectrometry of single tablets.
A near-IR camera using BEST image-analysis software permits multiple tablets to
be analyzed simultaneously in blister packages. Improvements in precision and
sample throughput may come with 1. placing more pixels on the samples (eg, using cameras with a higher density
of detectors in the focal plane arrays), 2. using multiple camera views and light
source positions at different angles to collect more diffuse reflectance, and 3.
increasing image integration time. Applications of near-IR spectrometry are published almost every day in the
pharmaceutical literature. Imaging and computing technology are changing many areas of
scientific research. The next major advance of near-IR spectrometry in pharmaceutical
analysis will likely come in the form of multispectral imaging, which enables large
numbers of samples to be analyzed simultaneously. 
| The authors would like to thank the National Science Foundation for its support of
this work through award number CHE-9257998. 
|
1.
Lodder RA, Hieftje GM. Analysis of intact tablets by near-infrared reflectance spectrometry. Appl Spectrosc. 1988;42:556-558.
2.
Lodder RA, Selby M, Hieftje GM. Detection of capsule tampering by near-infrared reflectance analysis. Anal Chem. 1987;59:1921-1930.
3.
Lodder RA, Hieftje GM. Detection of subpopulations in near-infrared reflectance analysis. Appl Spectrosc. 1988;42:1500-1512.
4.
Aldridge PK, Mushinsky RF, Andino MM, Evans CL. Appl Spectrosc. 1994;48:1272-1276.
5.
Drennen JK, Lodder RA. Nondestructive near-infrared analysis of intact tablets for determination of degradation product J Pharm Sci. 1990;79:622-627.
6.
Niemczyk TM, Delgado-Lopez MM, Allen FS. Quantitative determination of bucindolol concentration in intact gel capsules using Raman spectroscopy. Anal Chem. 1998;70(13):2762-2765.
7.
Dempsey RJ, Davis DG, Buice RG Jr, Lodder RA. Biological and medical applications of near-infrared spectrometry. Appl Spectrosc. 1996;50(2):18A-34A.
8.
Fischer M, Tran CD. Investigation of solid-phase peptide synthesis by the near-infrared multispectral imaging technique: a detection method for combinatorial chemistry. Anal Chem. 1999;71(13):2255-2261.
9.
Fischer M, Tran CD. Evidence for kinetic inhomogeneity in the curing of epoxy using the near-infrared multispectral imaging technique. Anal Chem. 1999;71(5):953-959.
10.
Tran CD, Cui Y, Smirnov S. Simultaneous multispectral imaging in the visible and near-infrared region: applications in document authentication and determination of chemical inhomogeneity of copolymers. Anal Chem. 1998;70(22):4701-4708.
11.
Ingle JD, Crouch SR. Spectrochemical analysis. Englewood Cliffs, NJ: Prentice-Hall, 1988.
12.
Isaksson T, Kowalski B. Piece-wise multiplicative scatter correction applied to near-infrared diffuse transmittance data from meat products. Appl Spectrosc. 1993;47(6):702-709.
13.
Dempsey RJ, Cassis LA, Davis DG, Lodder RA. Near infrared imaging and spectroscopy in stroke research: lipoprotein distributions and disease. Ann NY Acad Sci. 1997;820:149-169.
14.
Svensson O, Josefson M, Langkilde FW. Classification of chemically modified celluloses using a near-infrared spectrometer and soft independent modeling of class analogies. Appl Spectrosc. 1997;51(12):1826-1835.

|
|